Class I Recall of Draeger Medical, Inc. VentStar and ID Breathing Circuits and Anesthesia Sets

PUBLISHED: Jan 25, 2019
Relevant to: Ambulatory Care, Critical Access Hospitals, Hospitals, Long Term Care

The U.S Food and Drug Administration (FDA) is informing health care providers of the Class I recall of Draeger Medical, Inc. VentStar and ID Breathing Circuits and Anesthesia Sets. The FDA considers Class I Recalls the most serious type of recall. Use of these devices may cause serious injuries or death.

Recalled Product(s):

  • Draeger Medical, Inc. VentStar and ID Breathing Circuits and Anesthesia Sets
  • Model/Item Numbers:
  • MP00349, MP00350, MP00351, MP00352, MP00361, MP00374 (VentStar Circuits and Anesthesia Sets)
  • MP01341, MP01348, MP01350 (ID Circuits and Anesthesia Sets)
  • Manufacturing Dates: January 2016 to November 2018
  • Distribution Dates: April 2016 to December 2018
  • Devices Recalled in the U.S.: 1,200

Draegar Medical is recalling its disposable VentStar and ID Breathing Circuits and Anesthesia Sets due to a risk of the devices being incorrectly assembled, resulting in a short-circuit in the breathing hose. If the breathing hose is short circuited, the patient will not receive the expected breathing support (ventilation). Lack of breathing support may result in irreversible patient harm, up to and including severe oxygen-loss (hypoxia) and death.

On December 21, 2018, Draeger Medical sent customers an “Urgent Medical Device Recall Notification” and” Customer Reply and Order Form.” The recall notification instructed customers to:

  • Follow the device's Instructions for Use for correct setup of the breathing circuit, and check to ensure the connections are assembled correctly prior to each use.
  • If a short-circuited breathing circuit is not detected during a pre-use check, it will not be possible to ventilate the patient.
  • Check whether the breathing circuit is assembled so that the inspiratory and the expiratory connector of the ventilator or anesthesia machine are each connected to the y-piece before each use, and after any temporary disconnection.
  • Inspect your existing stock by following the provided inspection instructions and complete and return the “Customer Reply and Order Form” to confirm that inspections have been completed.
  • Use the “Customer Reply and Order Form” to order free replacements for any breathing circuits that were pre-assembled incorrectly.
  • An incorrectly pre-assembled breathing circuit can be detected because it will not be a single cohesive hose system but two separate sub-systems.

Customers who have questions or need additional information regarding this recall may contact Michael Kelhart between the hours of 8 AM – 4:30 PM EST at 1-800-437-2437 (press 1 at the prompt, then press 32349).

Follow the link below for additional information.