Medtronic Announces Voluntary Recall of its Pipeline Embolization Device, Alligator Retrieval Device, X-Celerator Hydrophilic Guidewire, Ultraflow and Marathon Flow Directed Micro Catheters

PUBLISHED: Oct 18, 2016

Medtronic is notifying customers of a voluntary recall of certain lots of its Pipeline™ embolization device, Alligator™ retrieval device and X-Celerator™ hydrophilic guidewire. The recall also includes the stylet containing UltraFlow™ flow directed micro catheters and Marathon™ flow directed micro catheters. These products are produced, marketed and sold by Medtronic's Neurovascular business, which is part of the Brain Therapies division in the company's Restorative Therapies Group.

This voluntary recall is being conducted due to the potential separation and detachment of the polytetrafluoroethylene (PTFE) coating on parts of these devices. Should the PTFE separate from the delivery wire or stylets, PTFE particulate could enter the blood stream of the patient. PTFE in the blood stream, based on the size and quantity, could lead to a thromboembolic event.

Medtronic initiated customer communication of the recall by letter on October 5, 2016, and is requesting customers to quarantine all affected product that remain in the inventory and return to Medtronic. The U.S. Food and Drug Administration (FDA) and other regulatory bodies also have been notified.

At the initiation of this recall, 84,278 units potentially affected by this recall had been distributed worldwide. The products were manufactured from July 2014 to September 2016. Additional information about the recall, including a link to a list of specific lot numbers of affected product, can be found by following the link below.

The Pipeline™ embolization device is indicated for the endovascular treatment of adults (22 years of age and older) with large or giant wide-necked intracranial aneurysms in the internal carotid artery from the petrous to the superior hypophyseal segments. The first generation Pipeline™ embolization device is affected by this action due to the PTFE coated delivery wire, which is part of the disposable delivery system (the braid implant is not affected). The second-generation device, Pipeline™ Flex embolization device, is not affected by this recall.

The Alligator™ retrieval device is intended for use in the peripheral and neuro-vasculature for foreign body retrieval. The X-Celerator™ hydrophilic guidewire is indicated for general intravascular use to aid in the selective placement of catheters in the peripheral, visceral and cerebral vasculature during diagnostic and/or therapeutic procedures.

The UltraFlow™ flow directed micro catheter is designed for the subselective infusion of physician-specified therapeutic agents such as embolization materials and diagnostic materials such as contrast media in tortuous, distal vessels.

The Marathon™ Flow Directed Micro Catheter is intended to access peripheral and neuro vasculature for the controlled selective infusion of physician-specified therapeutic agents such as embolization materials and of diagnostic materials such as contrast.